Toolbar - Calibration group
Tools specific to Devices tab:
Note: All these tools will eventually switch to Calibrations tab.
 
 
Calibrate
 
Import Calibration
 
Show Calibrations
Opens pop-up window with list of available calibration procedures. Clicking on Calibrate inside the pop-up will start external program (Caliber by default) to run this procedure.
Once the calibration is done, WinQBase will switch to Calibrations tab, use the data from external program to create new calibration record, associate it with selected device and activate Edit mode for the user fill in additional details.
 
Asks the user to import either data from previously created offline calibration or an external calibration report in PDF, XLS, XLSX, DOC or DOCX format. WinQBase will then switch to Calibrations tab, prompts the user to define calibration date, use the data to create new calibration record, associate it with selected device and activate Edit mode for the user fill in additional details.
Import of external reports automatically sets calibration status to external and further prompts the user to input certificate ID and issue date.
 
Switches to Calibrations tab and shows calibrations related to the selected device. This is done by filtering out all other calibrations from the List of Records.
 
 
 
 
 
 
 
 
Tools specific to Calibrations tab:
 
Redo Calibration
 
Standards List
Starts external program (Caliber by default) to re-run selected calibration.
Once the calibration ends, WinQBase will use the data from external program to create new calibration record, copy all other data from the previously selected calibration record and activate Edit mode for the user touch up any details as needed.
 
Opens pop-up window with list of standards used in selected calibration. The list can be printed in calibration report as required by ISO17025.
User may add and remove devices from current database flagged as Standards to/from the list using Add Standard and Remove Standard buttons in the pop-up window.